Table 1.

Suggested MRI protocol for wrist imaging at 3T (Wrist Coil)

Sequence FOV
(cm)
TR TE Slice
Thickness
(mm)*
Matrix
Cor T2 FS 8 5500 55 2.0 416×224
Ax T2 FS 8 5000 55 2.8 416×224
Cor PD FS 8 4000 35 2.0 384×192
Cor T1 8 750 20 2.0 416×192
Ax T1 8 750 20 2.8 416×192
Sag PD FS 8 4000 40 2.5 320×224
Cor CUBE Flex 12 2000 25 2 224×224
*

No interslice gap is utilized on any sequences, allowing for contiguous images.