. 2015 Jul 30;17(1):65. doi: 10.1186/s12968-015-0164-7

Table 4.

Effect of CMR scanner and acquisition protocol

archPWVWU Avg by 2 Avg by 3 Avg by 4
Linear fit slope Group 1 1.02 0.94 0.91
Group 2 1.05 0.96 0.88
Bland-Altman mean bias [limits] (m/s) Group 1 0.32 [-1.30 to 1.95] 0.30 [-1.54 to 2.16] 0.37 [-1.68 to 2.42]
Group 2 0.07 [-1.58 to 1.73] -0.06 [-2.51 to 2.38] -0.16 [-3.17 to 2.84]
Correlation r(p) Group 1 0.92 (<0.001) 0.90 (<0.001) 0.87 (<0.001)
Group 2 0.96 (<0.001) 0.90 (<0.001) 0.85 (<0.001)

For the wavelet-based method, aortic pulse wave velocity (archPWVWU) estimated from the original ascending and descending aortic flow curves, was used as a reference for comparisons against archPWVWU estimated from curves with lower temporal resolution obtained by averaging time points by blocks of 2 (Avg by 2), 3 (Avg by 3) and 4 (Avg by 4) in both Group 1 and Group 2