Table 2.
Material | In vitro CIC (μC.cm-2.ph-1) | In vivo CIC (μC.cm-2.ph-1) |
---|---|---|
PtIr | 23 5 | 18 5 |
TiN | 194 17 | 72 13 |
The CIC of each of the electrodes was measured using single biphasic symmetric current pulses with pulse widths of 100 s. The results are presented as mean SD.