Table 4.

Analytical results in μg L−1 for trace elements measured in seawater certified reference materials (mean ± one SD). LOD, limit of detection.

SLEW-3
CASS-4
Measured (n = 5) Certificate value Measured (n = 5) Certificate value
V 2.80 ± 0.06 2.57 ± 0.31 1.25 ± 0.11 1.11 ± 0.16
Mn 1.67 ± 0.04 1.61 ± 0.22 2.92 ± 0.01 2.78 ± 0.19
Ni 1.37 ± 0.06 1.23 ± 0.07 0.329 ± 0.043 0.314 ± 0.030
Co 0.033 ± 0.003 0.042 ± 0.010 0.033 ± 0.004 0.026 ± 0.003
Cu 1.63 ± 0.04 1.55 ± 0.12 0.644 ± 0.042 0.592 ± 0.055
Zn <0.159 (LOD) 0.201 ± 0.037 0.498 ± 0.175 0.381 ± 0.057
As 1.04 ± 0.10 1.36 ± 0.09 1.28 ± 0.14 1.11 ± 0.16
Cd 0.044 ± 0.008 0.048 ± 0.004 0.020 ± 0.006 0.026 ± 0.003
Pb <0.017 (LOD) 0.0090 ± 0.0014 <0.017 (LOD) 0.0098 ± 0.0036