Table 1.

Minimum SMBG System Accuracy Criteria According to the ISO 15197 Standard in Its 2003 and 2013 Versions and the FDA Draft Guidance Document From 2014.

ISO 15197:2003 ISO 15197:2013 FDAa
Relative number of results 95% 95%b 95% 99%
Within ±15 mg/dl ±20% ±15 mg/dl ±15% ±15% ±20%
At BG concentrations <75 mg/dl ≥75 mg/dl <100 mg/dl ≥100 mg/dl Entire range
99% of results within CEG zones A + B
a

FDA draft guidance for Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, published in 2014.

b

In ISO 15197:2013 this acceptance criterion is also applied for the user performance evaluation.