Skip to main content
. 2015 Apr 14;9(4):885–894. doi: 10.1177/1932296815580160

Table 3.

User Performance Evaluation Following ISO 15197 Standard.

ISO 15197:2003 ISO 15197:2013
Acceptance criteria 95% of measurements obtained within ±15 mg/dl of the reference measurement results at glucose concentrations <100 mg/dl and within ±15% at glucose concentrations ≥100 mg/dl
Number of test strip lots 3 1
Performance of measurements with the SMBG system At least 50 subjects At least 100 diabetic subjects
Sample type Capillary whole blood Capillary whole blood
Determination of accuracy Results measured by the subjects shall be compared to Results measured by the subjects shall be compared to
• results obtained with the reference measurement procedure results obtained with the reference measurement procedure
• results obtained by medical/laboratory personnel from the same sample
Reference measurement procedure Manufacturer’s measurement procedure Any method with verified metrological traceabilitya
Collection of samples for reference measurements Within 5 minutes after the subject’s measurement with the SMBG system by medical/laboratory personnel Within 5 minutes after the subject’s measurement with the SMBG system by medical/laboratory personnel
Training material/practice tests Instructions for use, etc routinely provided with the system Instructions for use, etc routinely provided with the system/subjects may be allowed a limited number of practice tests
Incorrectly performed measurements The measurement can be repeated if the subject reports a mistake (a maximum of 3 data exclusions due to incorrectly performed measurements shall be allowed)
Human factors User technique observation by the medical/laboratory personnel User technique observation by the medical/laboratory personnel
Evaluation of the instruction for use Ease of understandingb Ease of understandingc
a

Not clearly defined but indicated.

b

The evaluation method is not described in detail.

c

The evaluation method is described in detail.