Table 2.

Parameters Referring to Clamp Quality and Utility

Clamp no. Trueness of BG measurement (mg/dl) CV at target level (%) MARD at target level (%) Mean diff. device vs target level (mg/dl) Utility (%)
1 2.3 3.7 5.2 0.4 97.0
2 9.3 7.0 13.6 0.4 96.9
3 8.8 5.7 8.6 0.0 84.3
4 0.9 4.2 5.7 0.3 99.7
5 3.6 3.5 2.2 0.4 98.0
6 −0.1 4.7 4.5 0.7 96.7
7 −0.9 11.0 10.9 0.0 98.8
8 −4.3 4.3 6.6 1.2 100.0
9 4.7 5.1 6.7 1.0 95.0
10 2.4 9.7 8.9 0.1 98.7
11 0.8 5.6 4.6 1.8 90.3
12 1.0 2.6 5.4 0.1 98.5
13 0.9 6.2 4.5 2.5 98.4
14 0.7 4.9 4.5 0.3 96.6
15 −0.2 7.7 5.6 2.1 96.8
16
17 2.4 5.3 5.9 0.2 98.0
18 0.4 9.4 7.5 2.0 94.5
19 0.0 4.2 3.8 0.8 99.2
20 2.1 5.2 5.5 0.3 95.8
21 0.2 8.4 3.8 8.6 91.3
22 1.2 5.5 4.0 1.5 93.7
23 1.6 3.7 4.3 1.1 97.9
24 1.9 5.4 4.6 0.3 91.9
25 −0.1 4.3 3.3 1.4 98.3
26 0.4 8.5 4.4 3.3 83.5
27 1.7 3.0 3.3 0.5 94.5
28 1.3 2.3 4.9 0.4 97.9
29 2.9 6.1 4.5 0.1 94.9
30 −1.1 7.6 6.4 1.8 88.5
31
32 1.2 7.8 3.4 3.6 98.6
33 −1.1 6.2 4.9 2.5 89.5
34 3.3 6.6 4.9 1.9 94.7
35 −1.0 4.5 3.6 0.6 98.9
36 −1.5 3.8 3.2 1.3 91.2
37 −0.3 4.9 7.5 0.5 97.9
38 −0.3 3.5 3.6 0.9 98.6
39 −1.4 2.4 2.3 0.3 95.8
Mean 1.2 5.5 5.3 1.2 95.4
SD 2.5 2.1 2.3 1.6 4.1
Median 0.9 5.2 4.6 0.7 96.8
Interquartile range 2.3 -2.4 2.1 1.5 3.9

Trueness: mean (device vs ref. method); CV: 100 × (SD device/mean device); MARD: 100 × ([ref. meth. – device]/ref. meth.); utility: 100 × (operational time/total time).