Table 2.
Clamp no. | Trueness of BG measurement (mg/dl) | CV at target level (%) | MARD at target level (%) | Mean diff. device vs target level (mg/dl) | Utility (%) |
---|---|---|---|---|---|
1 | 2.3 | 3.7 | 5.2 | 0.4 | 97.0 |
2 | 9.3 | 7.0 | 13.6 | 0.4 | 96.9 |
3 | 8.8 | 5.7 | 8.6 | 0.0 | 84.3 |
4 | 0.9 | 4.2 | 5.7 | 0.3 | 99.7 |
5 | 3.6 | 3.5 | 2.2 | 0.4 | 98.0 |
6 | −0.1 | 4.7 | 4.5 | 0.7 | 96.7 |
7 | −0.9 | 11.0 | 10.9 | 0.0 | 98.8 |
8 | −4.3 | 4.3 | 6.6 | 1.2 | 100.0 |
9 | 4.7 | 5.1 | 6.7 | 1.0 | 95.0 |
10 | 2.4 | 9.7 | 8.9 | 0.1 | 98.7 |
11 | 0.8 | 5.6 | 4.6 | 1.8 | 90.3 |
12 | 1.0 | 2.6 | 5.4 | 0.1 | 98.5 |
13 | 0.9 | 6.2 | 4.5 | 2.5 | 98.4 |
14 | 0.7 | 4.9 | 4.5 | 0.3 | 96.6 |
15 | −0.2 | 7.7 | 5.6 | 2.1 | 96.8 |
16 | — | — | — | — | — |
17 | 2.4 | 5.3 | 5.9 | 0.2 | 98.0 |
18 | 0.4 | 9.4 | 7.5 | 2.0 | 94.5 |
19 | 0.0 | 4.2 | 3.8 | 0.8 | 99.2 |
20 | 2.1 | 5.2 | 5.5 | 0.3 | 95.8 |
21 | 0.2 | 8.4 | 3.8 | 8.6 | 91.3 |
22 | 1.2 | 5.5 | 4.0 | 1.5 | 93.7 |
23 | 1.6 | 3.7 | 4.3 | 1.1 | 97.9 |
24 | 1.9 | 5.4 | 4.6 | 0.3 | 91.9 |
25 | −0.1 | 4.3 | 3.3 | 1.4 | 98.3 |
26 | 0.4 | 8.5 | 4.4 | 3.3 | 83.5 |
27 | 1.7 | 3.0 | 3.3 | 0.5 | 94.5 |
28 | 1.3 | 2.3 | 4.9 | 0.4 | 97.9 |
29 | 2.9 | 6.1 | 4.5 | 0.1 | 94.9 |
30 | −1.1 | 7.6 | 6.4 | 1.8 | 88.5 |
31 | — | — | — | — | — |
32 | 1.2 | 7.8 | 3.4 | 3.6 | 98.6 |
33 | −1.1 | 6.2 | 4.9 | 2.5 | 89.5 |
34 | 3.3 | 6.6 | 4.9 | 1.9 | 94.7 |
35 | −1.0 | 4.5 | 3.6 | 0.6 | 98.9 |
36 | −1.5 | 3.8 | 3.2 | 1.3 | 91.2 |
37 | −0.3 | 4.9 | 7.5 | 0.5 | 97.9 |
38 | −0.3 | 3.5 | 3.6 | 0.9 | 98.6 |
39 | −1.4 | 2.4 | 2.3 | 0.3 | 95.8 |
Mean | 1.2 | 5.5 | 5.3 | 1.2 | 95.4 |
SD | 2.5 | 2.1 | 2.3 | 1.6 | 4.1 |
Median | 0.9 | 5.2 | 4.6 | 0.7 | 96.8 |
Interquartile range | 2.3 | -2.4 | 2.1 | 1.5 | 3.9 |
Trueness: mean (device vs ref. method); CV: 100 × (SD device/mean device); MARD: 100 × ([ref. meth. – device]/ref. meth.); utility: 100 × (operational time/total time).