Table 3.

FDA-Cleared BGM Systems Meeting ISO 15197 2003 and ISO 15197 2013 Standards.

Meter Name Company ISO 15197 2003 (32/33 evaluated) ISO 15197 2013 (31/33 evaluated)
Accu-Chek Active Roche Diagnostics Yes Yes
Accu-Chek Aviva Roche Diagnostics Yes Noa
Accu-Chek Aviva Plus Roche Diagnostics Yes Yes
Accu-Chek Compact Roche Diagnostics Yes Yes
Accu-Chek Compact Plus Roche Diagnostics Yes Yes
Accu-Chek Go Roche Diagnostics Nob Nob
Accu-Chek Performa (maltose dependent) Roche Diagnostics Yes Yes
Accu-Chek Performa (maltose independent) Roche Diagnostics Yes Yes
Advocate Redi-Code TaiDoc Technology Corp No No
alphacheck professional i-SENS inc Yes Yes
Ascensia Contour Bayer HealthCare LLC Yes ?
BG Star AgaMatrix Yes Noc
Breeze Bayer Health Care ? No
CareSens N i-SENS inc Yes Yes
CareSens N POP i-SENS inc Yes Yes
Clever Chek TD-4222 Taidoc Technology Corp. No No
Contour Bayer Consumer Care Nob No
Contour USB Bayer Consumer Care AG Yes Nob
Contour USB Next Bayer Health Care Yes Yes
Element Infopia Co Yes No
EZ Smart Tyson Bioresearch No No
FreeStyle Freedom Abbott Diabetes Care Inc Yes ?
FreeStyle Lite Abbott Diabetes Care Inc Yes Nod
GM700 Bionime Yes Yes
iBG Star AgaMatrix Yes Noe
microdot Cambridge Sensors Limited Yes No
OneTouch Select Lifescan No No
OneTouch Ultra 2 LifeScan Inc Yes Yes
OneTouch Verio Lifescan Inc Yes Yes
OneTouch VITA Lifescan Inc Yes Yes
Prodigy Voice Diagnostic Devices Inc No No
TRUEbalance Nipro Diagnostics No No
WaveSense Jazz AgaMatrix Yes Yes
FDA-cleared BGM systems meeting standards when tested for ISO standards 24/32 (75%) 15/31 (48%)

Question mark indicates the BGM system was not evaluated against this standard. Of the 98 BGM systems, 33 were FDA-cleared. Of those, 32 were evaluated for ISO 15197 2003 and 31 were evaluated for ISO 15197 2013.

a

Did not meet criteria in 1 of 5 studies.

b

Did not meet criteria in 1 of 2 studies.

c

Did not meet criteria in 1 of 4 studies.

d

Did not meet criteria in 1 of 4 lots of strips in a single study.

e

Did not meet criteria in 1 of 3 studies.