Table I.

Summary of measured ultrasound safety parameters of the four detection pulses

Single pulse Barker 13 Chirp short Chirp long FDA Regulatory Limit
MI 0.44 0.47 0.54 0.61 1.90
TI 0.01 0.04 0.08 0.16 6
ISPTA, 0.3 (mW/cm2) at 2500 Hz PRF 34.4 137.3 153.7 321.6 720
TR (°C) Probe surface Within tissue < 0.1 < 0.1 < 0.1 0.2 6
0.008 0.032 0.036 0.075