Skip to main content
. 2015 Aug;28(8):989–998. doi: 10.1016/j.echo.2015.02.020

Table 3.

Test-retest reproducibility results for 2D KBR and 2D echocardiographic RV metrics

Variable Intraobserver
Interobserver
ICC COV (%) RD (%) ICC COV (%) RD (%)
2D KBR
 RV EDV 0.985 3.0 4.2 0.986 3.9 5.5
 RV ESV 0.987 4.3 6.1 0.960 7.7 10.9
 RV SV 0.953 8.3 11.7 0.856 11.7 16.5
 RV EF 0.919 6.4 9.0 0.758 10.5 14.8
2DE
 RV EDA 0.885 9.0 12.7 0.394 25.0 35.4
 RV ESA 0.931 9.3 13.2 0.440 31.0 43.9
 RV FAC 0.784 18.1 25.6 0.619 20.8 29.4

COV, Coefficient of variation; EDA, end-diastolic area; EDV, end-diastolic volume; EF, ejection fraction; ESA, end-systolic area; ESV, end-systolic volume; ICC, intraclass correlation coefficient; RD, relative difference; SV, stroke volume.