Skip to main content
. 2015 Feb 2;17(8):1282–1288. doi: 10.1093/europace/euu339

Table 3.

Results for the CARTO pacing study according to anatomical site paced

Heart chamber RA (N = 92) LA (N = 167) RV (N = 61) LV (N = 92) RA + LA + RV + LV (N = 412)
Mean distance ± SD (mm) 7.4 ± 2.7 6.9 ± 2.3 6.5 ± 2.1 6.4 ± 2.2 6.8 ± 2.4

Mean distance between reference pacing site and reconstructed pacing site and standard deviation for N pacing events.

RA, right atrium; LA, left atrium; RV, right ventricle; LV, left ventricle; SD, standard deviation.