Skip to main content
. Author manuscript; available in PMC: 2015 Aug 14.
Published in final edited form as: Curr Protoc Cytom. 2009 Jul;0 12:Unit–12.17. doi: 10.1002/0471142956.cy1217s49

Table 12.17.1.

Lens Selections for Optical Apparatus shown in Figure 12.17.1

Field of view Camera lens Dichroic to array distance Distance between dichroic boxes Focal lens of secondary lenses
4 × 4 mm2 Nikon 50 mm f/1.2a 75 mm 65 mm No lens
5 × 5 mm2 Nikon 85 mm f/1.4 75 mm 85 mm No lens
1 × 1 cm2 Nikon 85 mm f/1.4 75 mm 30 mm 100 mm
2 × 2 cm2 Nikon 85 mm f/1.4 75 mm 30 mm 75 mm
a

An alternative is to use an inverted 25-mm Navitar lens.