Table I. Use of the smaller dimension pediatric oxygenator resulted in greater negative pressure.

Pressure readings were measured pre and post oxygenator in mmHg. All circuits were circulated at a flow rate of 1 liter/minute. Median values (25th, 75th percentile) are displayed. N=4.

Table 1A
PEDIATRIC ADULT p-value
Premembrane 126 (122,131) 121 (117, 124) 0.09
Postmembrane 110 (104,134) 118 (113, 120) 0.05
Pressure Gradient 16 (15,21) 3 (3,4.5) <0.05
Revolutions per minute 1497 (1468,1521) 1394 (1379,1420) 0.18
Table 1B
ADULT/HF ADULT p-value
Premembrane 115 (107,137) 121 (117, 124) 0.42
Postmembrane 113 (105,134) 118 (113, 120) 0.23
Pressure Gradient 2 (1,4) 3 (3,4.5) 0.01
Revolutions per minute 1366 (1307,1510) 1394 (1379,1420) 0.67