Table 4.
Specimen | Maximal load [N] | Vertical stiffness KV [N/mm] | Lateral stiffness KL [N/mm] | Number of cycles | Failure types |
---|---|---|---|---|---|
TomoFix std 3 | 1280 | 1350 | 2000 | >60 000 | 2 |
TomoFix std 4 | 1440 | 2000 | 2500 | >80 000 | 2 |
TomoFix std 5 | 1760 | 2500 | 2200 | >120 000 | 2 |
PEEKPower 3 | 1440 | 2000 | 2500 | >80 000 | 2 |
PEEKPower 4 | 1280 | 1950 | 2140 | >60 000 | 2 |
PEEKPower 5 | 1440 | 2785 | 2250 | >80 000 | 2 |
iBalance 3 | 1760 | 4000 | 3600 | >120 000 | 2,4 |
iBalance 4 | 1760 | 3000 | 3400 | >120 000 | 2 |
iBalance 5 | 1920 | 3000 | 2952 | >140 000 | 2 |
iBalance 6 | 1760 | 3500 | 2500 | >120 000 | 1,2 |