Table 4.

IL-1β ELISA data obtained from dendrimer plate, kit plate, by physical adsorption.

concentration [pg ml-1] Dendrimer Plate [O.D.] IL-1β kit [O.D.] Physical Adsorption [O.D.]
250 1.781 1.806 1.215
62.5 0.519 0.561 0.712
15.6 0.146 0.149 0.576
3.9 0.055 0.043 0.512
1.95 0.044 0.030 0.506
0.975 0.034 - 0.499
0 0.021 0.014 0.491

R2 0.996 0.995 0.991

LLD 0.04 0.032 0.516

Sensitivity [pg ml-1] <1.15 <2.08 <3.93