Skip to main content
. 2015 Aug 18;53(9):2970–2976. doi: 10.1128/JCM.01431-15

TABLE 3.

Correlation between LAMP and real-time PCR for M. pneumoniae detection from specimen extracts

LAMP result No. (%) of specimens
qPCR CT < 30 (n = 153) qPCR CT ≥ 30 (n = 73) qPCR negative (n = 26) Total
Positive 152 (99.3) 48 (65.8) 1 (0.04) 201 (79.8)
Negative 1 (0.7) 25 (34.2) 25 (96.2) 51 (20.2)
Total 226 (89.7) 26 (10.3) 252 (100)