Table 2.

Contrast recovery and background variability parameters when applying CT-based AC

Sphere size [mm] Radioactive / Non-radioactive aContrast recovery [%] aBackground variability [%]
10 Radioactive 30.5 ± 1.3 3.8 ± 0.8
13 Radioactive 50.5 ± 2.4 3.0 ± 0.5
17 Radioactive 72.9 ± 1.9 2.5 ± 0.2
22 Radioactive 74.5 ± 3.3 2.3 ± 0.2
28 Non-radioactive 56.6 ± 2.3 2.0 ± 0.2
37 Non-radioactive 64.8 ± 0.9 1.8 ± 0.1

a The values represent the mean values and standard deviation of four independent measurements for the reconstruction: 3 iterations, 21 subsets, 172 matrix and 4 mm Gaussian filtering