Published in final edited form as: Ther Drug Monit. 2015 Oct;37(5):661–669. doi: 10.1097/FTD.0000000000000201

Table 3.

Intra-plate imprecision (n=7) for JWH-018 N-(5-hydroxypentyl) concentrations (1–10 µg/L). Inter-plate imprecision (n=35) for JWH-018 metabolite using positive controls (10 & 20µg/L) and performance challenge concentrations at ±25% and ±50% of each cutoff. Inter-read imprecision for a representative calibration curve (1–500µg/L) for JWH-018 N-(5-hydroxypentyl) in urine.

Urine JWH-018
metabolite µg/L
Intra-plate (n=7)
Absorbance %
Inter-plate (n=35)
Absorbance %
Inter-read (n=7)
Absorbance %



Mean (SD) CV Mean (SD) CV Mean (SD) CV
1 0.827 (0.051) 6.2 - - 0.902 (0.024) 2.7
2.5 0.613 (0.027) 4.2 0.787 (0.092) 11.6 0.727 (0.020) 2.8
3.75 - - 0.630 (0.079) 12.5 - -
5 0.511 (0.036) 7.0 0.594 (0.067) 11.2 0.474 (0.015) 3.2
6.25 - - 0.557 (0.067) 12.0 - -
7.5 0.469 (0.038) 8.2 0.555 (0.078) 14.0 0.407 (0.014) 3.5
10 0.327 (0.020) 5.5 0.388 (0.035) 9.0 0.374 (0.011) 3.1
12.5 - - 0.351 (0.035) 10.0 - -
15 - - 0.321 (0.031) 9.8 - -
20 - - 0.284 (0.037) 13.0 - -
25 - - - - 0.222 (0.007) 3.3
50 - - - - 0.157 (0.006) 3.7
100 - - - - 0.118 (0.005) 4.0
250 - - - - 0.087 (0.003) 3.7
500 - - - - 0.070 (0.001) 2.1