Skip to main content
. 2015 Aug 2;2015:853506. doi: 10.1155/2015/853506

Table 1.

Sequence parameters for liver 3.0-T MR imaging in rats with HCC.

Precontrast T1-weighted T2-weighted Diffusion weighted Postcontrast T1-weighted (3 phases)
Sequence FSPGR SSFSE SE/EPI FSPGR
Repetition time (msec) 200 5000 6000 200
Echo time (msec) 2.1 83.6 Minimal 2.1
Flip angle (degree) 70 NA NA 70
Matrix 192 × 256 192 × 256 64 × 64 192 × 256
Field of view (cm2) 10 × 7 10 × 7 10 × 7 10 × 7
Section thickness (mm) 3 3 3 3
Intersection gap (mm) 0.3 0.3 0.3 0.3
Number of excitations 6 1 4 6
Number of slices 13 13 13 13
b-value used (sec/mm2) NA NA 0, 800 NA

FSPGR: fast spoiled gradient-recalled echo, SSFSE: single shot fast spin-echo, Se/EPI: spin-echo/echo-planar, TR: repetition time, E: echo time, NA: not applicable.