Published in final edited form as: IEEE Trans Biomed Eng. 2015 Mar 26;62(9):2177–2186. doi: 10.1109/TBME.2015.2415731

TABLE II.

SUMMARY OF ISV SURFACE RECONSTRUCTION AND PATIENT REGISTRATION ACCURACIES USING LANDMARK AND ISV TECHNIQUES. THE COMPUTATIONAL COST FOR MULTI-START ISV REGISTRATION IS ALSO INCLUDED.

Pat. # iSV reconstruction Registration accuracy (mm)3 iSV registration cost (sec)
# of probe data Accuracy (mm)2 Landmark iSV-based
1 4 2.50±1.12 1.95±1.25 1.27±0.91 33
2 5 2.22±1.03 1.83±1.50 1.09±0.73 94
3 4 2.38±0.99 1.93±1.30 1.20±0.77 134
4 5 1.73±0.44 1.73±0.11 1.11±0.61 106
5 8 2.64±1.12 2.71±1.80 1.90±1.41 73
6 5 1.94±0.89 2.01±1.62 1.91±1.82 100
7 4 2.45±0.51 1.98±1.03 1.70±1.12 42
8 6 2.10±1.00 2.03±1.7 1.30±1.02 184
Avg. 51 2.21±0.31 2.02±0.30 1.43±0.35 95.8±48.9
1

Median value;

2

iSV reconstruction accuracy is the average point-wise distance error relative to the probe points;

3

Registration accuracy is the average SDE relative to pCT