Skip to main content
. 2014 May 1;14(5):1–163.

Table 13:

Identifying IV Infusions: Performance Metrics and Analysis

Performance Metrics and Analysis Pump Identification Error Line-Tracing Error
Infusion Disconnection Taska Infusion Disconnection Taska Infusion Documentation Task
Definition Stop wrong pump Disconnect wrong IV tubing (of identified pump) Identify wrong access port
      Identify wrong infusion (x 3) connected to identified access port
Performance metric (per participant per condition) Pass or fail Pass or fail Number of line-tracing errors (maximum of 4)b
Analysis
Number of participants included in analysis 40 40 39c
Opportunities for error per condition 40 (1 error per disconnection task; 1 disconnection task per participant per condition) 40 (1 error per disconnection task; 1 disconnection tasks per participant per condition) 156 (4 errors per documentation task; 1 documentation task per participant per condition)
Statistical test (performance metric as a function of experimental condition) Cochran's Q test followed by pairwise comparisons between the different combinations of intervention conditions by use of the McNemar χ2 test with Bonferroni correction NA (statistical analysis not conducted, since no errors were made across all conditions) One-way ANOVA test followed by post hoc paired sample t-test comparisons using Bonferroni correction

Abbreviations: ANOVA, analysis of variance; IV, intravenous; NA, not applicable.

a

In the infusion disconnection task, there was the potential for 2 errors: identifying/stopping the wrong pump and tracing the infusion to the wrong patient access port/disconnecting the wrong infusion.

b

In the infusion documentation task, there was the potential for 4 line-tracing errors: identifying the wrong access port and connected infusions (3 times).

c

One participant (of 40) could not complete the documentation task with the light-linking system because of technical difficulties.