Fig. 3.
Examples of CARTO maps of a repeat case. In the default setting (a) and setting #1 (b), the anterior wall is not shown as a red area. However, in the optimized setting #2 (c) the anterior wall is shown in red. The abated areas are large, but there are few ablated points excluding the red areas. Thus, setting #2 is useful for excluding the areas that do not require ablation.