Table 2.

Variance of kmax over the entire mesh surface. Results for Gynemesh (n=5) and Restorelle (n=5) samples are shown for both 1 and 10N under boundary conditions (BCs) 1, 2, and 3. Reported p-values are from a Friedman’s test, which tested the impact of BCs on Var(kmax).

Var(kmax) (cm−1)
Mesh Load
(N)
BC 1 BC 2 BC 3 p
Gynemesh 1 0.006
(0.0034 - 0.0169)
0.0279
(0.0236 - 0.084)
0.148
(0.124 - 0.240)
0.041
10 0.05*
(0.0438 - 0.0534)
0.4721
(0.2341 - 0.7265)
5.714*
(3.386 - 6.341)
0.007
Restorelle 1 0.006
(0.0045 - 0.0077)
0.047
(0.0422 - 0.0705)
0.0851
(0.0707 - 0.127)
0.015
10 0.0139*
(0.0119 - 0.0143)
0.458
(0.248 - 0.537)
1.855*
(1.774 - 2.787)
0.007
*