Skip to main content
. 2014 Nov 12;50(2):493–518. doi: 10.1007/s10853-014-8685-2

Table 1.

Si(Li)-EDS analysis of NIST Standard Reference Material 482 Copper–Gold Alloysa (all concentration values in mass fraction)

Cu (certified) Analyzed Rel. error (%) Au (certified) Analyzed Rel. error (%) Total
0.198 0.198 ± 0.002 0.0 0.801 0.790 ± 0.002 −1.4 0.988
0.396 0.399 ± 0.001 0.8 0.603 0.594 ± 0.002 −1.6 0.993
0.599 0.605 ± 0.001 1.0 0.401 0.402 ± 0.002 0.1 1.007
0.798 0.797 ± 0.001 −0.1 0.200 0.199 ± 0.003 −1.2 0.996

Analysis performed with Cu Kα and AuLα; beam energy = 20 keV; quantitative calculations with NIST Desktop Spectrum Analyzer [12]; uncertainty expressed as 1σ from the measured counts (from Ref [13])

aSee footnote 2