Skip to main content
. 2015 Aug;5(4):503–510. doi: 10.3978/j.issn.2223-4292.2015.04.07

Table 1. Effect of heart rate on MOLLI performance vs. reference T1 SE in phantom gels.

Scheme Co-efficient of variability
HR 40 HR 60 HR 80 HR 100
Flip angle 35°
   5,3,0 6.77 7.61 9.49 7.57
   3,3,5 6.81 7.21 8.31 8.85
   4,3,2 6.94 7.78 7.89 9.55
Flip angle 50°
   5,3,0 7.86 7.71 7.60 9.19
   3,3,5 7.74 7.99 8.67 11.10
   4,3,2 9.00 9.30 9.03 8.89

MOLLI, modified look-locker inversion recovery; HR, heart rate.