TABLE II.

Ecorr, Icorr and Epitt Readings Determined from Potentiodynamic Polarization Curves for Plasma Coatings Deposited onto 316L Stainless Steel Stents in PBS Electrolyte with Ag/AgCl Reference Electrode

Treatment Ecorr (mV) Icorr (μA/cm2) Epitt (mV)
316L SS –198.72 ± 22.53 4.09 ± 2.39 +461.4 ± 43.1
DC-TMS +76.95 ± 6.93* 0.63 ± 0.10 +693.3 ± 27.7*
DC-TMS + NH3/O2 +9.88 ± 34.37* 2.54 ± 1.99 +710.6 ± 9.4*
RF-TMS –64.58 ± 51.08* 0.16 ± 0.05 +612.0 ± 17.4*
RF-TMS + NH3/O2 –14.87 ± 50.60* 0.12 ± 0.08 +562.0 ± 67.6*

The data are presented as means 6 standard deviations for n = 3.

*

(denotes p< 0.05, as compared to 316L SS).