Skip to main content
. Author manuscript; available in PMC: 2015 Sep 4.
Published in final edited form as: Anal Methods. 2014 Feb 25;6(8):2592–2607. doi: 10.1039/C3AY42085J

Table 5.

Fluorescent readings of separate samples of OPA in DMSO-reagent solution and blanks

Time elapsed after
addition of OPA
Fluorescence reading of
samplea
Fraction of maximum
reading
Fluorescence reading
of blank
Fraction of maximum
readingb
20 min 1747 84.1% 4.951 7.94%
40 min 1781 85.7% 8.860 14.2%
60 min 1891 91.0% 13.89 22.3%
80 min 1924 92.6% 17.31 27.8%
100 min 1965 94.6% 20.46 32.8%
2 h 1974 95.0% 23.91 38.4%
3 h 2059 99.1% 31.00 49.7%
4 h 2056 98.8% 32.61 52.3%
5 h 2078 100.0% 38.46 61.7%
6 h 2022 97.3% 41.56 66.7%
7 h 1994 96.0% 43.42 69.7%
23 h 1270 61.1% 60.60 97.2%
26 h 1046 50.3% 62.32 100.0%
a

Fluorescence readings of samples were not corrected for blank readings.

b

Maximum reading of blank for this 26 h period.