Table 2.

Phantom composition and settings used in the CDMAM measurements

Equivalent breast thickness (mm) Total PMMA thickness (mm) Phantom composition* (mm) FFDM LE CESM
Target Filter Tube voltage (kV) Tube load (mAs) AGD (mGy) Target Filter Tube voltage (kV) Tube load (mAs) AGD (mGy)
32 30 10/Ph/10 W Rh 26 61 0.79 Mo Rh 27 45 1.1
45 40 10/Ph/10/10 W Rh 28 86 1.1 Rh Rh 29 71 1.8
60 50 10/10/Ph/10/10 W Rh 32 81 1.3 Rh Rh 31 80 2.3
75 60 25/Ph/25 W Ag 30 111 1.7 Rh Rh 30 140 3.6
90 70 25/Ph/25/10 W Ag 34 96 1.8 Rh Rh 30 140 3.8

*Ph = CDMAM phantom, 10 and 25 refer to the thickness of the PMMA plates (mm)

FFDM full-field digital mammography, LE low energy, CESM contrast-enhanced spectral mammography, AGD average glandular dose