Skip to main content
. 2015 Sep 9;42(10):5745–5756. doi: 10.1118/1.4929978

TABLE II.

Statistics of local dose difference comparison at 5 cm depth between MC simulations and 2D array measurements for US probes in horizontal orientation.

X6-1 C5-2
6 MV 15 MV 6 MV 15 MV
Mean (%) 0.3 −0.3 0.1 0.0
Standard deviation (%) 1.3 1.4 1.2 1.2
Max (%) 5.3 3.6 5.9 4.3
Voxels with <3% local difference 95.7 96.6 97.4 99.1