Skip to main content
. 2015 Sep 11;11:43. doi: 10.1186/s13007-015-0083-5

Table 1.

Instruments and sensors for measurement and their calibration

What to measure Units Measured by Precision of instrumentb Accuracy of readingb Calibrated by and when
Radiation (PARa) µmol m−2 s−1 Quantum sensor ±1 % ±10 % Comparison with a reference sensor or against a standard quartz-halogen lamp traceable to a (inter)national standard (e.g. US NIST). Once per annum
Radiation (net) W m−2 Net radiometer ±2 % ±5 % Comparison with a reference meter or in a temperature controlled calibration chamber capable of evaluating short and long wavebands. Once per annum
Radiation (spectral) µmol m−2 s−1 nm−1 Spectroradiometer ±1 % ±5 % Comparison with a reference meter or against a standard source of radiation (e.g. ASTM G138-06). Once per annum
Irradiance (solar) W m−2 Pyranometer ±1 % ±5 % Comparison with a reference meter or against a standard source (traceable to the World Radiometric Reference) in an integrating sphere. Once per annum
Radiation (integral) MJ m−2 d−1 or mol m−2 d−1 Calculated from accumulated radiation data
Air temperature °C RTD, thermocouple or thermistor (shaded and aspirated in air with speed ≥3 m s−1) ±0.1 °C ±0.2 °C Comparison with a reference thermometer (e.g. traceable against US NIST) by placing in melting crushed ice and boiling distilled water. Once per annum
Substrate temperature °C RTD, thermocouple or thermistor (ensure good contact with substrate) ±0.1 °C ±0.2 °C As above
Surface temperature °C Infrared temperature sensor, fine wire thermocouple ±0.1 °C ±0.2 °C Infrared sensor: Comparison with a reference surface thermometer mounted on the same surface within the field of view of the infrared sensor
Thermocouple: as above
Atmospheric moisture: relative humidity or vapour pressure deficit (VPD) % or kPa Capacitance, dewpoint sensor, psychrometer, or IRGA (infrared gas analyser) Relative humidity: ±2 %
Dewpoint temp.: ±0.1 °C
VPD: ±0.3 kPa
±5 %
±0.5 °C
±0.5 kPa
Humidity generator; unsaturated salt solution calibration standards (35 and 80% RH). Once per annum
Air speed m s−1 Anemometer (range 0.1–15.0 m s−1) ±2 % ±5 % Wind tunnel. Once per annum
pH pH probe (range 3–10) ±0.1 pH ±0.1 pH Standard solutions. Before every measurement, or weekly in continuous measurement applications
Electrical conductivity (EC) S m−1 Electrical conductivity meter ±3 % ±5 % As above
Dissolved oxygen mg L−1 Dissolved oxygen meter (maintain adequate solution flow rate and ensure temperature compensation) ±3 % ±5 % As above
Atmospheric CO2 concentration µmol mol−1 Silicon based NDIR (non-dispersive infrared) sensor as part of an IRGA (infrared gas analyser) ±1 % ±3 % Certified calibration gases for low and high end of the measurement range, and/or precision gas mixing instrument. Once per week

aReferred to as photosynthetically active radiation (PAR: 400–700 nm) for general usage and described as photosynthetic photon flux density (PPFD) by many journals, professional societies and manufacturers of quantum sensors.

bPrecision is how close the measured values are to each other. Accuracy is how close a measured value is to the actual (true) value.