Published in final edited form as: J Immunol Methods. 2014 May 22;408:78–88. doi: 10.1016/j.jim.2014.05.008

Table 2.

Assessment of assay accuracy and precision for detecting PA83 and PA63. Assay was repeated in duplicate 10 times (n) and for each the number of runs with a measurable results (nm), expected mean (μM) the observed mean (μM), standard deviation (St. dev), coefficient of variation (CV), and percent error (% error).a

Toxin Expected mean
(μM)
n m /n Observed
mean (μM)
St. dev CV
(%)
% Error
PA83 0 3/10 0.0000028 0.0000023 81.7 ND
0.000002 6/10 0.0000019 0.0000011 55.9 4.7
0.00001 9/10 0.000014 0.000026 189.7 37.7
0.001 10/10 0.0009 0.0002 20.8 11.8
0.005 10/10 0.0041 0.0013 31.8 18.4
0.01 10/10 0.012 0.005 41.2 23.5
0.02 9/10 0.051 0.04 77.2 156.2
PA63 0 0/10 ND ND ND ND
0.000009 0/10 ND ND ND ND
0.00003 3/10 0.00012 0.00002 133.2 58.4
0.008 10/10 0.0057 0.0005 8.2 23.4
0.02 10/10 0.022 0.005 23.9 9.0
0.05 10/10 0.055 0.018 32.7 10.5
0.2 3/10 0.15 0.04 27.0 24.5
a

ND = not determined