Skip to main content
. Author manuscript; available in PMC: 2016 Oct 7.
Published in final edited form as: Phys Med Biol. 2015 Sep 15;60(19):7499–7512. doi: 10.1088/0031-9155/60/19/7499

Table 1.

Independently measured (provided by the manufacturer) and estimated attenuations of five CIRS phantoms (unit: dB/cm/MHz)

Phantom No. Independently measured Test-1 Test-2 Test-3 Test-4 Test-5 Mean SD
1 0.28 0.176 0.193 0.224 0.255 0.221 0.214 0.031
2 0.57 0.573 0.503 0.495 0.424 0.435 0.486 0.06
3 0.90 0.78 0.787 0.618 0.58 0.659 0.685 0.944
4 1.17 0.857 0.776 0.834 0.869 0.67 0.801 0.082
5 1.45 1.092 0.974 0.95 0.852 0.836 0.941 0.104