Table 1.

Intra- and inter-assay precision and accuracy for IgG, IgM, and IgA

QC samples Intra-assay precision (CV%) Intra-assay accuracy (% bias) QC samples Inter-assay precision (CV%) Inter-assay accuracy (% bias)
IgG IgG
 LLOQ (3.2 ng/mL) 12.3 −17.1  LLOQ (3.9 ng/mL) 19.1 −16.5
 Low QC (7.2 ng/mL) 10.7 −16.7  LQC (8.6 ng/mL) 18.6 −7.7
 Middle QC (14.2 ng/mL) 9.5 −8.6  MQC (15.6 ng/mL) 19.9 −16.9
 High QC (33.5 ng/mL) 6.7 −2.3  HQC (34.3 ng/mL) 15.2 −8.7
 ULOQ (57.7 ng/mL) 17.1 −9.2  ULOQ (63.5 ng/mL) 16.5 −15.3
IgM IgM
 LLOQ (43.4 ng/mL) 7.1 11.8  LLOQ (49.3 ng/mL) 14.0 5.1
 Low QC (91.8 ng/mL) 4.4 −8.7  LQC (100.5 ng/mL) 7.5 7.2
 Middle QC (190.8 ng/mL) 4.2 −5.3  MQC (201.6 ng/mL) 5.9 7.5
 High QC (371.2 ng/mL) 3.6 −5.4  HQC (392.3 ng/mL) 6.6 4.6
 ULOQ (791.8 ng/mL) 4.3 −1.3  ULOQ (802.5 ng/mL) 5.9 7.0
IgA IgA
 LLOQ (4.0 ng/mL) 3.7 −9.3  LLOQ (4.5 ng/mL) 11.1 −4.9
 Low QC (8.5 ng/mL) 2.7 −5.9  Low QC (9.0 ng/mL) 7.0 −3.8
 Middle QC (17.0 ng/mL) 2.6 −3.9  Middle QC (17.7 ng/mL) 5.8 −5.5
 High QC (35.3 ng/mL) 2.8 −2.7  High QC (36.3 ng/mL) 4.9 −3.3
 ULOQ (71.9 ng/mL) 3.5 −1.5  ULOQ (73.0 ng/mL) 4.4 −2.6

CV coefficient of variation, LLOQ lower limit of quantitation, QC quality control, LQC low range of the dynamic range QC, MQC middle of the dynamic range QC, HQC high range of the dynamic range QC, QC quality control, ULOQ upper limit of quantitation