Skip to main content
. Author manuscript; available in PMC: 2015 Sep 18.
Published in final edited form as: Anal Methods. 2014 May 12;6(13):4698–4704. doi: 10.1039/C4AY00271G

Table 1.

Selected Ion Monitoring (SIM) parameters, limit of detection (LOD), and calibration curve range/linearity for the quantitation of ten flavor analytes.

SIM Ions, m/z
Compound Retention Time
(min)
Dwell time (ms) LOD
(μg/g)
Calibration Range
(μg/g)
Linearity, R2
(Average)
Quant. Ion Conf. Ion
Eucalyptol 2.79 154.2 (75) 139.1 (75) 5.69 5.02 – 10041 0.998
Camphor 3.16 152.1 (50) 108.1 (85) 3.69 4.86 – 9725 0.997
Menthol 3.20 138.2 (65) 123.1 (65) 5.07 5.04 – 10090 0.998
Methyl Salicylate 3.30 120.1 (65) 152.1 (65) 0.95 5.18 – 10356 0.999
Pulegone 3.47 152.1 (75) 137.1 (100) 3.12 4.91 – 9813 0.998
Ethyl Salicylate 3.58 166.1 (65) 120.0 (50) 0.44 5.02 – 10042 0.998
Cinnamaldehyde 3.59 131.1 (40) 103.1 (65) 1.08 5.07 – 10136 0.997
Eugenol 3.98 164.1 (55) 131.1 (75) 0.75 4.91 – 9822 0.999
Diphenyl Ether 4.32 170.1 (75) 141.1 (90) 0.28 5.03 – 10056 0.999
Coumarin 4.61 149.0 (50) 118.1 (50) 0.38 5.08 – 10160 0.999
MDA (ISTD) 4.53 164.1 (50) 146.0 (50)

ISTD = Internal Standard

R2 = Coefficient of Determination, Linearity