Skip to main content
. 2015 Aug 21;49(3):307–313. doi: 10.1515/raon-2015-0021

TABLE 3.

Passing rates of three consecutive measurement results using film, diode array (Mapcheck), ion chamber array (MatriXX) and electronic portal imaging device (EPID) based on gamma index values for 6 fields of patient 1. The data shown for patient 1 in Table 2 is the average of first measurement set of data in Table 3

Measurement Field 1 Field 2 Field 3 Field 4 Field 5 Field 6
Film 1st 98.95 98.97 91.20 97.86 98.88 98.66
2nd 97.90 98.93 91.18 98.74 98.28 98.71
3rd 97.58 98.42 91.15 99.24 98.65 98.26
Mean (SD) 98.14 (0.59) 98.77 (0.25) 91.18 (0.02) 98.61 (0.57) 98.60 (0.25) 98.54 (0.20)
Mapcheck 1st 100.00 100.00 100.00 100.00 100.00 100.00
2nd 100.00 100.00 100.00 100.00 100.00 100.00
3rd 100.00 100.00 100.00 100.00 100.00 100.00
Mean (SD) 100.00 (0.00) 100.00 (0.00) 100.00 (0.00) 100.00 (0.00) 100.00 (0.00) 100.00 (0.00)
MatriXX 1st 99.29 98.41 99.57 99.30 99.10 98.97
2nd 99.30 99.25 99.57 99.28 99.06 99.01
3rd 99.29 99.24 99.58 99.29 99.00 98.99
Mean (SD) 99.29 (0.00) 98.97 (0.39) 99.57 (0.00) 99.29 (0.01) 99.05 (0.04) 98.99 (0.02)
EPID 1st 99.80 99.40 99.60 98.90 98.80 100.00
2nd 99.90 99.60 99.70 99.20 99.10 100.00
3rd 99.80 99.60 99.45 99.05 98.90 99.90
Mean (SD) 99.83 (0.05) 99.53 (0.09) 99.58 (0.10) 99.05 (0.12) 98.93 (0.12) 99.97 (0.05)

SD = standard deviation