Skip to main content
. 2015 Aug 21;49(3):307–313. doi: 10.1515/raon-2015-0021

TABLE 4.

Passing rates based on gamma evaluation using 3 films in the same location for intensity modulated radiation therapy (IMRT) quality assurance (QA) of each patient

Patient 1 Patient 2 Patient 3 Patient 4 Patient 5
Film 98.36 95.29 98.51 98.08 99.41
97.47 95.47 98.55 98.19 99.14
97.63 95.65 98.87 98.36 99.27
Mean 97.82 95.47 98.64 98.21 99.27
SD 0.39 0.15 0.16 0.12 0.11