Table 4. The limits of detection for the Anyplex assay.

Concentration of mutant DNA (%)* Positives/total runs Mutation detection rate (%)
3.000 8/8 100
2.000 8/8 100
1.500 8/8 100
1.000 8/8 100
0.500 8/8 100
0.100 8/8 100
0.050 8/8 100
0.010 4/8 50
0.005 4/8 50
0.001 1/8 12.5

*The SNU-790 BRAF V600E-positive cell line was serially diluted with the DMPK-M BRAF V600E-negative cell line; The limit of detection claimed by the manufacturer was 1.000% of BRAF mutations. All results were positive when the concentration was 0.050%. The limit of detection as determined by probit analysis was 0.046% (95% confidence interval:0.019-0.532). Reproducibility was verified at 1.000% and lower concentrations.