Table 4. The limits of detection for the Anyplex assay.
| Concentration of mutant DNA (%)* | Positives/total runs | Mutation detection rate (%) |
|---|---|---|
| 3.000 | 8/8 | 100 |
| 2.000 | 8/8 | 100 |
| 1.500 | 8/8 | 100 |
| 1.000† | 8/8 | 100 |
| 0.500 | 8/8 | 100 |
| 0.100 | 8/8 | 100 |
| 0.050 | 8/8 | 100 |
| 0.010 | 4/8 | 50 |
| 0.005 | 4/8 | 50 |
| 0.001 | 1/8 | 12.5 |
*The SNU-790 BRAF V600E-positive cell line was serially diluted with the DMPK-M BRAF V600E-negative cell line; †The limit of detection claimed by the manufacturer was 1.000% of BRAF mutations. All results were positive when the concentration was 0.050%. The limit of detection as determined by probit analysis was 0.046% (95% confidence interval:0.019-0.532). Reproducibility was verified at 1.000% and lower concentrations.