Skip to main content
. 2015 Sep 14;65(5):293–300. doi: 10.1007/s13224-015-0774-x

Table 2.

Comparison of thermal damage at different power settings among different laparoscopic shears

Group comparison Mean value A (μm) Mean value B (μm) p value
Monopolar diathermy vs. Harmonic scalpel (output power 3) 215.79 90.42 <0.001
Monopolar diathermy vs. Harmonic scalpel (output power 5) 215.79 127.48 <0.001
Monopolar diathermy v. LigaSure 215.79 144.18 <0.001
Harmonic scalpel (output power 3) vs. Harmonic scalpel (output power 5) 90.42 127.48 0.001
Harmonic scalpel (output power 3) vs. LigaSure 90.42 144.18 <0.001
Harmonic scalpel (output power 5) vs. LigaSure 127.48 144.18 0.39
Harmonic scalpel (output power 5)vs. Sonosurg 90.49 117.6 <0.001