Negative testing results |
Decomposed reaction buffer and solutions |
Prepare new reaction buffer and solutions and protect from light |
|
Air bubbles in the wells of the black 96-well plate |
Tap the black 96-well plate before incubation in the multiplate reader |
|
Divergent testing results |
Incomplete lysis of the cells |
Prepare new lysis buffer and strongly pipette up and down when lyzing cells |
|
Variation in temperature |
Check temperature in the multiplate reader and ensure a constant temperature of 37°C |
|
Variation in pH and crystallization of the reaction buffer |
Check the pH of the reaction buffer and adjust to 7.4 |