. 2015 Oct;56(10):1926–1935. doi: 10.1194/jlr.M060343

TABLE 2.

Validation results of CT and 7-KC by HPLC-ESI-MS/MS

CT 7-KC
Calibration curve, n = 5
 Mean slope (range; SD) 0.0080 (0.0074–0.0089; 0.0006) 0.0044 (0.0037–0.0049; 0.0005)
 Mean intercept (range; SD) 0.39 (0.29–0.44; 0.061) 0.10 (0.09–0.12; 0.013)
 Mean coefficient of linear regression, r2 (range; SD) 0.9908 (0.9830–0.999; 0.006) 0.9926 (0.9817–0.9975; 0.0058)
Imprecision (% CV) Accuracy (%) Imprecision (% CV) Accuracy (%)
Within day, n = 10
 12.5 ng/ml 5.5 3.8 5.5 0.3
 100 ng/ml 3.6 1.7 4.4 3.7
 800 ng/ml 2.4 8.9 3.9 4.4
Between day, n = 10
 12.5 ng/ml 5.8 1.5 5.8 4
 100 ng/ml 5.8 5.3 7.9 6.4
 800 ng/ml 2.9 0.5 7.7 5.4
Recovery (%), n = 3
 25 ng/ml 94.4 94
 200 ng/ml 98.2 107
 400 ng/ml 91.8 100.7
LOD (ng/ml), n = 3 1 1
LOQ (ng/ml), n = 3 3 3