Skip to main content
. 2015 Sep 18;7:101–111. doi: 10.2147/NSS.S77888

Table 2.

Sensitivity and PPV of Z-PLUS as compared with the PSG Consensus for Wake, Light Sleep, Deep Sleep, and REM

Wake Light Sleep Deep Sleep REM
Sensitivity 0.91 0.84 0.74 0.72
PPV 0.84 0.85 0.78 0.73

Abbreviations: PPV, positive predictive value; PSG, polysomnography; REM, Rapid Eye Movement.