Table 2.
Proportion of patients with poor (<30%) LV capture safety margin (LVSM) with different pacing strategies (LV threshold +1V and triple PW at LV threshold, 0.4ms). PW, pulse width.
LV Pacing Configurations | Pacing output at LV threshold + 1 Volt % patients LVSM < 30% |
Pacing output at triple PW (1.2ms) at LV threshold % patients LVSM < 30% |
---|---|---|
| ||
Tip→RVcoil | 21% | 67% |
Tip→Can | 25% | 53% |
Tip→Ring | 42% | 61% |
Ring→RVcoil | 14% | 68% |
Ring→Can | 14% | 64% |
Ring→Tip | 42% | 68% |
Note: LV capture safety margin refers to the relationship between LV pacing output and LV threshold (see methods for details calculating LVSM).