Table 1. The standard procedures of IONM.
Procedures | Note |
---|---|
Record vocal cord movement before surgery | Use of fiberoptic laryngoscope |
Recommended anesthesia method for IONM | Intermediate- or short-acting muscle relaxant is recommended for preoperative anesthesia induction. The dosage of intermediate-acting muscle relaxant is smaller than that used for routine anesthesia induction. 1x ED95 intermediate-acting non-depolarizing muscle relaxant is recommended; avoid adding muscle relaxant before the termination of intraoperative monitoring |
Body position and tube placement | Anesthesia intubation is performed after positioning (intubation under video laryngoscope is recommended) |
Device connections and checking | |
The ground electrodes are routinely placed under the skin at the shoulders or xiphoid | |
Confirm the monitoring system has been effectively established | |
Check electrode impedance and differences in impedance values | Electrode impedance <5 kΩ, with deviations <1 kω |
Check initial EMG | Initial fluctuations: about 10 μV |
Set up event thresholds | Typically 100 μV |
The current intensity of stimulator probe should be routinely set at 1-3 mA | |
The monitoring device should be placed far away from electro-surgical devices and be connected with anti-jamming silence detectors | |
The recording electrode positions should be confirmed during surgery | Directly detect, locate, and record the electrode depth at the antemedial laryngeal line using stimulator probe |
IONM four-step method | |
Step 1: V1 signal | Obvious bipolar EMG signal is obtained at the ipsolateral vagus nerve at the plexas thyreoidea inferior level (point B), confirming the successful establishment of the monitoring system. If signal is absent at point B, detect the vagus nerve at the plexas thyroid superior level (point A); the presence of signal at point A confirms the presence of non-recurrent laryngeal nerve (25) (Figure 4) |
Step 2: R1 signal | Before the exposure of RLN, its EMG signal is located using the Cross method by applying the probe vertical to trachea at its traveling area and then parallel to trachea |
Step 3: R2 signal | Continuous monitoring is applied during the dissection of RLN and the signal change is compared in a real-time manner. After the RLN is exposed, the most proximal end of the exposed part is detected for EMG signal |
Step 4: V2 signal | After complete hemostasis is achieved at the surgical field, the EMG signal of the vagus nerve is detected before closing the incision |
Signal analysis | |
Basic EMG parameters | The biphasic waveform should be differentiated from the monophasic non-EMG artifacts |
No obvious decrease in R2 and V2 signals | The basic EMG parameters include amplitude, latency and duration (26) (Figure 5); the RLN has intact function |
Loss of R2 and V2 signals | If the RLN is injured during the surgical operation, detect the “injury site” and search for the injury cause* |
Photo recording the exposed RLN during surgery | To confirm the RLN continuity (visual integrity) |
Postoperative laryngoscopy |
*, If no “injury site” is detected, it is important to determine whether there is a “real” loss of signal. (I) stimulate the nerve and then observe the contact between endotracheal tube electrode and vocal cord via a laryngoscope; (II) detect signals at vagus nerve and RLN again before closing the incision. IONM, intraoperative neuromonitoring; RLN, recurrent laryngeal nerve.