Skip to main content
. 2015 Aug 31;15(8):e28895. doi: 10.5812/hepatmon.28895

Table 2. Inter-Assay Reproducibility of HCV Real-Time RT-PCR.

Day Input Copy Numbers of HCV 5’NCR RNA Control
Threshold Cycle (Ct)
103 104 105 106 107
1 29.21 26.47 22.06 19.88 15.28
2 29.50 26.61 23.30 19.23 15.91
3 30.10 26.03 21.54 19.06 15.12
4 29.43 25.40 22.21 19.41 14.79
5 28.65 27.23 20.87 20.37 16.32
Values a 29.39 ± 0.52 26.34 ± 0.68 21.99 ± 0.89 19.61 ± 0.50 15.48 ± 0.61
CV b 1.76 2.58 4.04 2.54 3.94

a Data are presented as mean ± SD.

b Abbreviation: coefficient of variation.