Table 1.
Measured system efficiency with USER operating at 1 MHz, for different transducer diameters, source power, and propagating medium.
Expt. # | Transmitter-Receiver transducer diameter (mm) | Medium | Pin (W) | Battery current (mA) | Efficiency (%) |
---|---|---|---|---|---|
1 | 25 | None, direct charging | 0.6 | 93 | 50 |
2 | 10 | 1-cm water | 4.0 | 29 | 3 |
3 | 25 | 1-cm water | 4.0 | 155 | 15 |
4 | 40 | 1-cm water | 4.0 | 208 | 20 |
5 | 25 | 10 mm porcine tissue | 4.0 | 150 | 15 |
6* | Ti+PZT 25* | 3–5 mm porcine tissue | 2.0 | 110 | 22 |
7* | Ti+PZT 25* | 6–10 mm porcine tissue | 3.0 | 67 | 10 |
8 | 25 | 50 mm of 10 mm slices | 7.0 | 20 | 1.1 |
9 | 25 | 50 mm beef muscle | 7.0 | 70 | 4 |
Experiments performed with receiver transducer bonded inside a titanium pacemaker casing.