Table 1.

Performance of the Hi3-1 Multiplex kit compared against the Architect system on samples of panel 1.

  Abbott diagnostics  
(i) HIV 1/2: Architect HIV Ag/Ab Combo test  
(ii) HCV: Architect anti-HCV
Concordance rate (%)
  Positive∗ Negative
Hi3-1
Multiplex HIV 1/2
     
 Positive 102 2 99.96
 Negative 0 4477
Hi3-1
Multiplex HCV
     
 Positive 431 10 99.78
 Negative 0 4140