Skip to main content
. 2015 Sep 17;2015:805367. doi: 10.1155/2015/805367

Table 2.

Bland-Altman plot comparing automated and manual central corneal thickness (aCCT, mCCT), anterior chamber depth (ACD), and nasal and temporal trabecular iris angle (nTIA, tTIA) measurements by swept-source optical coherence tomography SS OCT and time-domain optical coherence tomography TD OCT in control and corneal dystrophies group (CD group). 95% CI:95% confidence interval of the mean difference; SD: standard deviation.

Control group CD group
Mean difference 95% CI SD p Mean difference 95% CI SD p
aCCT −4.4 −10.47–1.63 21.29 .14 −4.32 −8.51–0.13 20.35 .04
mCCT −3.4 −7.7–0.97 15.37 .12 −3.75 −8.81–1.31 24.58 .14
ACD 0.07 0.06–0.09 0.06 <.001 0.06 −0.01–0.14 0.38 .01
nTIA −1.58 −2.48–−0.67 3.2 .001 1.8 0.14–3.47 6.62 .03
tTIA −2.22 −3.04–−0.39 2.91 <.001 −1.97 −2.59–−1.36 2.97 <.001