Skip to main content
. Author manuscript; available in PMC: 2015 Oct 1.
Published in final edited form as: Anal Methods. 2012 Jan 26;4(2):513–521. doi: 10.1039/C2AY05686K

Table 1.

Γ in different buffers

Buffer ± SD (μm) # of beads Γ ± SDa (μm3 s–2)
Sizing 0.456 ± 0.015 337 102 ± 3
1.025 ± 0.01 318 110 ± 1
1.898 ± 0.029 368 109 ± 2
Mean 107 ± 3
Shrinking 0.456 ± 0.015 419 108 ± 3
1.025 ± 0.01 104 109 ± 1
1.898 ± 0.029 151 111 ± 2
Mean 109 ± 1
Swelling 0.456 ± 0.015 261 142 ± 5
1.025 ± 0.01 171 139 ± 1
1.898 ± 0.029 120 139 ± 2
Mean 140 ± 1
Combination 0.974 ± 0.034 69 98 ± 4
a

The Γ standard deviation (SD) for each trial includes the bead size uncertainty and the statistical uncertainty due to the number of beads studied.