Table 1. Sensitivity, Specificity, and Accuracy of PDFF Determined With MRI in the Detection of Clinically Relevant Steatosis.

Sensitivity [95% CI] Specificity [95% CI] Accuracy [95% CI]
Breath-hold 94 [73–100]% (17/18) 91 [75–98]% (29/32) 92 [81–98]% (46/50)
Bellows 94 [73–100]% (17/18) 90 [71–96]% (28/32) 90 [78–97]% (45/50)
Navigator 100 [81–100]% (18/18) 91 [75–98]% (29/32) 94 [83–99]% (47/50)

Numbers in parentheses are numbers of patients. Sensitivity, specificity, and accuracy of MRI fat fraction for the detection of steatosis was calculated using a reference threshold fat fraction of 5.56%, based on MRS PDFF as the reference standard.