Table 1.

Feature comparison across three femtosecond laser platforms.

Laser platform Catalys Victus LDV Z8
Company Abbott/Optimedica (USA) Bausch and Lomb (USA) Ziemer (Switzerland)

FDA approval Arcuate incision; anterior capsulotomy; lens fragmentation Arcuate incision; anterior capsulotomy; lens fragmentation; corneal flaps Pending

CE Mark Arcuate incision; anterior capsulotomy; lens fragmentation Arcuate incision; anterior capsulotomy; lens fragmentation; corneal flap Arcuate incision; anterior capsulotomy; lens fragmentation

Pulse frequency
 Lens fragmentation 120 KHz 80 KHz 2 MHz
 Anterior capsulotomy 120 KHz 80 KHz 1 MHz

Pulse duration 600 fs 400–550 fs Undisclosed

Pulse energy
 Lens fragmentation 3–30 µJ 7.0 µJ <50 nJ
 Anterior capsulotomy 3–30 µJ 6.8 µJ <50 nJ

Patient-laser interface Liquid interface (nonapplanating) Curved lens applanation, with ocular surface bathed in saline solution Liquid interface (nonapplanating)

Docking Vacuum docking Vacuum docking Vacuum docking

IOP during suction 28.9 mmHg
(baseline: 11.4 mmHg)
[44];
25.9 mmHg
(baseline: 10.3 mmHg)
[33]
42.1 mmHg
(baseline: 25.0 mmHg)
[35]
Unknown

Lens fragmentation safety margins 0.2 mm–1 mm (adjustable)
0.5 mm (default)
0.7 mm–1 mm (adjustable) 0.4 mm–1.5 mm (adjustable)
0.6 mm (default)

Lens fragmentation patterns Grid-pattern; radial (quadrants, sextants, and octants); grid + radial Concentric ring-like; radial (quadrants, sextants, and octants) Radial (default: sextants)

Capsulotomy incision depth 0.6 mm Unspecified 0.8 mm

Can create corneal flaps? No Yes Yes

Imaging 3D Fourier-domain OCT 3D Fourier-domain OCT, real-time 3D Fourier-domain OCT

Integrated bed Yes Yes No